PROVOX NID 17 FR, 8 MM 7102

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-12-02 for PROVOX NID 17 FR, 8 MM 7102 manufactured by Atos Medical Ab.

Event Text Entries

[32784592] Product have not been returned, examination is not possible. It is indicated that the prosthesis have been exposed to excessive force since the esophagus flange was ruptured and continuous insertions with a gel cap by the patient. In the provox capsule ifu it is stated that the product is only to be used by clinicians (other gelcaps are not allowed to be used).
Patient Sequence No: 1, Text Type: N, H10


[32784593] According to the complaint description; "patient reported that voice prosthesis extruded at home, she had voice prosthesis but continued coughing. She suctioned herself and noted that she suctioned the central dark blue stabilizing piece of the prosthesis from her trachea. When the voice prosthesis was brought into the clinic it was noted that the esophageal flange was torn along its attachment to the voice prosthesis shaft. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8032044-2015-00010
MDR Report Key5260227
Date Received2015-12-02
Date of Report2015-12-02
Date of Event2015-10-20
Date Mfgr Received2015-10-27
Date Added to Maude2015-12-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS MARIE LANGE
Manufacturer StreetPO-BOX 183 KRAFTGATAN 8
Manufacturer CityHORBY, SE-242 22
Manufacturer CountrySW
Manufacturer PostalSE-242 22
Manufacturer Phone415 19800
Manufacturer G1ATOS MEDICAL AB
Manufacturer StreetPO BOX 183 KRAFTGATAN 8
Manufacturer CityHORBY, SE-242 22
Manufacturer CountrySW
Manufacturer Postal CodeSE-242 22
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePROVOX NID 17 FR, 8 MM
Generic NameVOICE PROSTHESIS, LARYNGEAL (TAUB)
Product CodeEWL
Date Received2015-12-02
Catalog Number7102
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerATOS MEDICAL AB
Manufacturer AddressPO BOX 183 KRAFTGATAN 8 HORBY, SE-242 22 SW SE-242 22


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2015-12-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.