RELIANCE ENDOSCOPE PROCESSOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-12-03 for RELIANCE ENDOSCOPE PROCESSOR manufactured by Steris Canada Corporation.

Event Text Entries

[32794297] A steris service technician inspected the unit and found that water was leaking from the drip pan. The technician also found that the solenoid valve required replacement. The technician cleaned the strainer for the drip pan, replaced the solenoid valve, and then confirmed the unit to be operating according to specification.
Patient Sequence No: 1, Text Type: N, H10


[32794298] The user facility reported their reliance endoscope processing system was leaking water. No injury or procedural delay was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9680353-2015-00086
MDR Report Key5264595
Date Received2015-12-03
Date of Report2015-12-03
Date of Event2015-11-09
Date Mfgr Received2015-11-09
Date Added to Maude2015-12-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC 8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameRELIANCE ENDOSCOPE PROCESSOR
Generic NameENDOSCOPE PROCESSOR
Product CodeNZA
Date Received2015-12-03
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC 8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-03