CHROMID(TM) CPS?ELITE 416172

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2015-12-11 for CHROMID(TM) CPS?ELITE 416172 manufactured by Biomerieux Sa.

Event Text Entries

[33265980] Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[33265981] A customer in (b)(6) reported to biomerieux misdentification of bacteria when using product chromid(tm) cpslite translucent. A pink enterobacter clocae was present from a urine sample after incubation 18-24h 37? Initially it was noted as e. Coli however; indole testing was negative. The vitekidentification (id) returned a 97% id for e. Clocae. There was a delay reporting results however, there was no return of incorrect results and no patient harm reported.
Patient Sequence No: 1, Text Type: D, B5


[35029413] Biom? Rieux internal investigation was conducted. As defined in the package insert, chromid tm cps elite translucent agar is used for identification of the bacteria most commonly isolated in urinary tract infections:? E. Coli: spontaneous red to burgundy coloration of strains producing ss-glucuronidase (ss-gur) and/or ssgalactosidase (ss-gal). ? Enterococcus: spontaneous turquoise coloration of strains producing ss-glucosidase (ss-glu). ? Kesc: spontaneous blue to green coloration of strains producing ss-glucosidase (ss-glu). ? Proteeae: spontaneous diffuse brown coloration of strains producing deaminase. The vitek? 2 report for the submitted strain indicates this enterobacter expresses ss-gal and therefore, according to the instructions for use of the media, it is expected to have coloration observed by the customer. The enzymatic profile of this strain is not typical. The chromid tm cps elite translucent agar product limitations indicate certain species may produce colonies of the same characteristic color as e. Coli (ex. : citrobacter). The chromid tm cps elite agar is working in accordance with specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2015-00165
MDR Report Key5288980
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2015-12-11
Date of Report2015-11-24
Date Mfgr Received2015-11-24
Device Manufacturer Date2015-09-28
Date Added to Maude2015-12-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID(TM) CPS?ELITE
Generic NameCHROMID(TM) CPS?ELITE
Product CodeJXA
Date Received2015-12-11
Catalog Number416172
Lot Number1004327280
Device Expiration Date2016-01-24
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.