ED PULSECHECK 5.4

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-12-11 for ED PULSECHECK 5.4 manufactured by Picis Clinical Solutions, Inc..

Event Text Entries

[33963145] After investigation picis clinical solutions has confirmed the reported defect and will be providing a solution ina future ed pulse check software solution. A capa has been initiated to further analyze the root cause(s) of this defect and to implement corrective action as necessary. No adverse events have been attributed to this issue. Clinical review suggests there is a potential for ordering and administering a medication that could interact with a patient's current medications without a notification being provided due to the omission from their current medication list. The nature and extent of any untoward outcome would be directly related to the severity of the interaction between the medication ordered and the home medication that is omitted. The root cause for clinicians being unable to apply a medication to a patient visit has been identified and a software solution is being developed for this specific client, with plans to distribute to the ed pulse check client base as part of a future service pack as a preventive measure to remove risk on similar interfaces that might be deployed at other client sites in the future.
Patient Sequence No: 1, Text Type: N, H10


[33963146] Picis clinical solutions received a customer complaint on (b)(6) 2015: when a current medication is sent on the inbound adt via a08, and the user clicks 'n/a' initially, that medication is not able to be re-added either through the "search" or quicklist to the current visit. During the initial patient triage interview the nurse marks a specific medication that has come in from previous/external via adt as 'na'. So aspirin 81 mg po, for example, is marked n/a because the patient claims they are not taking that anymore. Sometime later in the visit, the pharmacist is reviewing the record and determines that aspirin should in fact be active as a current medication. That specific medication is not able to be added back to the current meds visit. It allows you to select it, but when you find and submit, the screen refreshes and nothing gets entered. The clinician is able to enter the medication as free text, however since the medication is not entered as structured, the ed pulse check application will not perform medication allergy interaction checking. It was determined that this issue is likely contained to one existing client on ed pulse check v5. 4 due to their interface design and configuration.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005244943-2015-00005
MDR Report Key5290728
Date Received2015-12-11
Date of Report2015-11-18
Date Mfgr Received2015-11-18
Date Added to Maude2015-12-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KERRY ARMSTRONG
Manufacturer Street100 QUANNAPOWITT PARKWAY SUITE 405
Manufacturer CityWAKEFIELD MA 01880
Manufacturer CountryUS
Manufacturer Postal01880
Manufacturer Phone7815573000
Manufacturer G1PICIS CLINICAL SOLUTIONS, INC.
Manufacturer Street100 QUANNAPOWITT PARKWAY SUITE 405
Manufacturer CityWAKEFIELD MA 01880
Manufacturer CountryUS
Manufacturer Postal Code01880
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameED PULSECHECK
Generic NameSOFTWARE, TRANSMISSION AND STORAGE
Product CodeNSX
Date Received2015-12-11
Model Number5.4
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPICIS CLINICAL SOLUTIONS, INC.
Manufacturer Address100 QUANNAPOWITT PARKWAY SUITE 405 WAKEFIELD MA 01880 US 01880


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.