[33429139] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[33429140] The customer received false negative benzodiazepine plus results in urine that were confirmed to be positive with lc/ms. No specific data or date of event was provided. It was unknown if any erroneous result was reported outside of the laboratory. There was no adverse event. The analyzer involved in the event was a cobas integra 400 plus serial number (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[35777496] Data was provided for five patient samples. It was found the calibration for lot 608892 had a lower curve than the previous lot. According to the customer, the qc results were good but the patient results were significantly lower.
Patient Sequence No: 1, Text Type: N, H10


[37070795] A specific root cause could not be determined. Retention material of the same lot of reagent was tested and the results were within specification.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2015-06028
MDR Report Key5294606
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2015-12-14
Date of Report2016-01-27
Date of Event2015-11-27
Date Mfgr Received2015-11-27
Date Added to Maude2015-12-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS BENZODIAZEPINES
Generic NameENZYME IMMUNOASSAY, BENZODIAZIPINE
Product CodeJXM
Date Received2015-12-14
Model NumberNA
Catalog Number04780248190
Lot Number608892
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-14

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.