TI OBA PLATE ANCHOR SCREW SELF-DRILLING 6MM 04.500.026.01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2015-12-15 for TI OBA PLATE ANCHOR SCREW SELF-DRILLING 6MM 04.500.026.01 manufactured by Synthes Usa.

Event Text Entries

[33582726] Device was used for treatment, not diagnosis. Unknown when device malfunctioned. (b)(4). Device is expected to be returned for manufacturer review/investigation, but has yet to be received. Subject device has not been received. Without a lot number, the device history record review and the investigation could not be completed; no conclusion could be drawn, as no product was received if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[33582727] It was reported that a surgeon performed a bone-anchored maxillary protraction (bamp) procedure and implanted orthodontic bone anchor (oba) implants. The patient came in for follow-up approximately one month ago. The surgeon noticed one of the oba implants was loose. The surgeon removed loose hardware on (b)(6) 2015 and put a new plate in. The patient is doing well. No surgical delay reported. This report is 1 of 4 for (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[35113593] Date returned to manufacturer. Subject device has been received and is currently in the evaluation process. Investigation is ongoing; no conclusion could be drawn as product is entering the complaint system. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[36498131] A product investigation was completed: the following device(s) was received: anchor t-plate, 4 hole, domed head, orthodontic bone anchor system (part 04. 500. 013, lot 7798054) and 1. 55mm midface self-drilling screw (part 04. 500. 026. 01, lot unknown, quantity: 3). The returned devices are in good condition. Both returned plates have been contoured and cut, with minor cosmetic damage which indicates that the plates were once implanted. The returned screws are in good condition with no functional damage and only minor marring around the hex drive. Due to the nature of the complaint, the complaint condition was unable to be replicated or verified. A visual inspection, complaint history review, drawing review, and risk assessment review were performed as part of this investigation. During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition. A visual inspection, complaint history review, drawing review, and risk assessment review were performed as part of this investigation. No product design issues or discrepancies were observed. This complaint is unconfirmed. The relevant drawings for the device(s) were reviewed. No drawing issues or discrepancies were noted. The design was determined to be suitable for the intended use when employed and maintained as recommended. Proper use and maintenance for the device(s) are addressed in the technique guide. The root cause of the complaint condition is unknown. No manufacturing or design issues were noted during the investigation. The design determined to be adequate for its intended use when used and maintained as recommended (b)(4) lot unknown. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2520274-2015-17883
MDR Report Key5299436
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2015-12-15
Date of Report2015-11-24
Date Mfgr Received2016-01-11
Date Added to Maude2015-12-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLINDA PLEWS
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone6107195000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTI OBA PLATE ANCHOR SCREW SELF-DRILLING 6MM
Generic NameIMPLANT, ENDOSSEOUS, ORTHODONTIC
Product CodeOAT
Date Received2015-12-15
Returned To Mfg2015-12-15
Catalog Number04.500.026.01
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSYNTHES USA
Manufacturer Address1302 WRIGHTS LANE EAST WEST CHESTER PA 19380 US 19380


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2015-12-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.