STATIC AND DYNAMIC COMPLIANCE 22 E 777404-101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-12-16 for STATIC AND DYNAMIC COMPLIANCE 22 E 777404-101 manufactured by Carefusion.

Event Text Entries

[34329553] (b)(4). The customer replaced the screw of the arm bracket which resolved the reported issue. The device is currently functioning without any further issues. The screws will be returned to carefusion for an evaluation. When we receive the results of the investigation a follow up will be submitted. Carefusion continues to track and trend any incident related to this issue. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[34329554] The customer reported that the arm of the pulmonary function testing (pft) machine fell off of the cart. He stated that the screws on the mounting bracket are stripped, got loose and fell off. The incident happened while testing a patient. The customer caught the arm as it was falling and there was no harm done to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2015-02478
MDR Report Key5304348
Date Received2015-12-16
Date of Report2015-11-23
Date of Event2015-11-23
Date Mfgr Received2015-11-23
Device Manufacturer Date2014-01-07
Date Added to Maude2015-12-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRESPIRATORY THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJILL RITTORNO
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer Phone8473628056
Manufacturer G1CAREFUSION
Manufacturer Street1100 BIRD CENTER DRIVE
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2015-12-16
Model Number22 E
Catalog Number777404-101
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-16

© 2026 FDA.report