ID-MTS READER SA MTS9945

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2004-09-10 for ID-MTS READER SA MTS9945 manufactured by Micro Typing Systems, Inc..

Event Text Entries

[37184322] This customer reported the lid of the reader sa was not staying open by itself. Customer did not claim alleged death or serious injury. On august 10, 2004, fe performed the following repairs: inspected the hinge and found the pin that connects the lid to the lid hinge mechanism shaft was missing and replaced the part. Fe confirmed lid was working properly after repairs, holding its position mid close and without dropping. Repairs returned equipment to expected operation. No injury occurred or was reported. In a letter issued march 4, 2004, customers were informed of the issue and asked to check the lid hinge mechanism and contact customer technical support if any issues were noted. Customers were also informed that field engineers would be performing additional actions at future service calls. These actions include inspection, lubrication and repair as necessary, of any issues with the lid hinge mechanism. Complaint (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[37184323] Lid on reader sa could slam shut if hinge assembly is broken. No death or serious injury was associated with this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1058138-2004-00035
MDR Report Key5305870
Date Received2004-09-10
Date of Report2004-09-10
Date of Event2004-08-09
Date Mfgr Received2004-08-09
Device Manufacturer Date2002-11-01
Date Added to Maude2015-12-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCATHERINE MCMULLEN, SITE DIRECTOR
Manufacturer Street1295 SW 29TH AVE
Manufacturer CityPOMPANO BEACH FL 33069
Manufacturer CountryUS
Manufacturer Postal33069
Manufacturer Phone9546239537
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameID-MTS READER SA
Generic NameBLOOD GROUPING CENTRIFUGE AND READER
Product CodeKSO
Date Received2004-09-10
Model NumberMTS9945
Catalog NumberMTS READER SA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMICRO TYPING SYSTEMS, INC.
Manufacturer Address1295 SW 29TH AVE POMPANO BEACH FL 33069 US 33069


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2004-09-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.