PIP SZ. 30 PROXIMAL PIP-200-30P-WW

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-12-21 for PIP SZ. 30 PROXIMAL PIP-200-30P-WW manufactured by Ascension Orthopedics.

Event Text Entries

[34761802] It was reported one of the implants broke when it was put in the patient and had to be taken out. The surgeon then proceeded with the same two implants and finished the case successfully. It was reported that although the device was in contact with the patient there was no patient injury.
Patient Sequence No: 1, Text Type: D, B5


[40740814] Integra has completed their internal investigation on 14jan2016. The investigation activities included: methods: evaluation of actual device. Review of device history records. Review of complaint history. Results: the review of manufacturing records showed no evidence of nonconformance that could have caused or contributed to the reported defect. A review of complaints for pip implant fractures (both intra-op and post-op) during the last 5 years showed (b)(4) complaints. During this period of time, there have been approximately(b)(4) units sold. The resulting overall rate of complaints is (b)(4)). Trend analysis of the yearly complaints does not represent an adverse trend. Conclusion: the cause of the failure has been identified as resulting from impacting the unsupported head when the stem of the implant has not been fully inserted into the medullary canal due to an improperly prepared oblique osteotomy.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1651501-2015-00044
MDR Report Key5313617
Date Received2015-12-21
Date of Report2015-12-03
Date Mfgr Received2016-01-14
Device Manufacturer Date2013-12-22
Date Added to Maude2015-12-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER MARIA LEONARD
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1ASCENSION ORTHOPEDICS
Manufacturer Street8700 CAMERON ROAD #100
Manufacturer CityAUSTIN TX 78754
Manufacturer CountryUS
Manufacturer Postal Code78754
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePIP SZ. 30 PROXIMAL
Generic NamePIP
Product CodeNEG
Date Received2015-12-21
Returned To Mfg2016-01-08
Catalog NumberPIP-200-30P-WW
Lot Number132871T
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerASCENSION ORTHOPEDICS
Manufacturer Address8700 CAMERON ROAD #100 8700 CAMERON ROAD #100 AUSTIN TX 78754 US 78754


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.