MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2015-12-21 for NAVIOS FLOW CYTOMETER SYSTEM A52103 manufactured by Beckman Coulter Ireland.
[34015141]
A field service engineer (fse) evaluated the instrument. The fse found that the lifter motor had failed the fse replaced the motor, which repaired the failing vortex. The repairs were verified by the fse. The beckman coulter internal identifier for this event is (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[34015142]
The customer reported the vortex on the navios instrument was not working. Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
Patient Sequence No: 1, Text Type: D, B5
[38905364]
The beckman coulter internal identifier for this event is (b)(4). The beckman coulter internal identifier was changed from (b)(4) to the correct identifier, (b)(4).
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1061932-2015-01856 |
MDR Report Key | 5314271 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2015-12-21 |
Date of Report | 2015-11-23 |
Date of Event | 2015-11-23 |
Date Mfgr Received | 2015-11-23 |
Device Manufacturer Date | 2009-12-01 |
Date Added to Maude | 2015-12-21 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. MIRANDA HOLLAND |
Manufacturer Street | 11800 SW 147TH AVE 32-L23 |
Manufacturer City | MIAMI FL 33196 |
Manufacturer Country | US |
Manufacturer Postal | 33196 |
Manufacturer Phone | 3053802031 |
Manufacturer G1 | BECKMAN COULTER |
Manufacturer Street | 11800 SW 147TH AVE |
Manufacturer City | MIAMI FL 33196 |
Manufacturer Country | US |
Manufacturer Postal Code | 33196 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | NAVIOS FLOW CYTOMETER SYSTEM |
Generic Name | FLOW CYTOMETRIC REAGENTS AND ACCESSORIES |
Product Code | OYE |
Date Received | 2015-12-21 |
Model Number | NA |
Catalog Number | A52103 |
Lot Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BECKMAN COULTER IRELAND |
Manufacturer Address | 250 SOUTH KRAEMER BLVD. LISMEEHAN O'CALLAGHAN'S MILLS COUNTY CLARE, EI |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2015-12-21 |