NAVIOS FLOW CYTOMETER SYSTEM A52103

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2015-12-21 for NAVIOS FLOW CYTOMETER SYSTEM A52103 manufactured by Beckman Coulter Ireland.

Event Text Entries

[34015141] A field service engineer (fse) evaluated the instrument. The fse found that the lifter motor had failed the fse replaced the motor, which repaired the failing vortex. The repairs were verified by the fse. The beckman coulter internal identifier for this event is (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[34015142] The customer reported the vortex on the navios instrument was not working. Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
Patient Sequence No: 1, Text Type: D, B5


[38905364] The beckman coulter internal identifier for this event is (b)(4). The beckman coulter internal identifier was changed from (b)(4) to the correct identifier, (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2015-01856
MDR Report Key5314271
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2015-12-21
Date of Report2015-11-23
Date of Event2015-11-23
Date Mfgr Received2015-11-23
Device Manufacturer Date2009-12-01
Date Added to Maude2015-12-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MIRANDA HOLLAND
Manufacturer Street11800 SW 147TH AVE 32-L23
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal33196
Manufacturer Phone3053802031
Manufacturer G1BECKMAN COULTER
Manufacturer Street11800 SW 147TH AVE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameNAVIOS FLOW CYTOMETER SYSTEM
Generic NameFLOW CYTOMETRIC REAGENTS AND ACCESSORIES
Product CodeOYE
Date Received2015-12-21
Model NumberNA
Catalog NumberA52103
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER IRELAND
Manufacturer Address250 SOUTH KRAEMER BLVD. LISMEEHAN O'CALLAGHAN'S MILLS COUNTY CLARE, EI


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-21