VITROS CHEMISTY PRODUCTS AMON SLIDES 1726926

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-12-21 for VITROS CHEMISTY PRODUCTS AMON SLIDES 1726926 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[34763107] The investigation determined that a non-reproducible lower than expected vitros amon quality control result was obtained from one level of a vitros control fluid using vitros amon slides on a vitros 5600 integrated chemistry system. A definitive assignable cause for the event was not identified. As limited information was available, and the customer did not want to continue the investigation, a vitros amon reagent issue and a vitros instrument malfunction cannot be ruled out as contributing factors.
Patient Sequence No: 1, Text Type: N, H10


[34763108] The customer observed a non-reproducible lower than expected vitros amon quality control result from one level of a non - vitros control fluid using vitros amon slides on a vitros 5600 integrated chemistry system. Vitros amon result: 187. 41 ug/dl vs. Expected 234. 8 ug/dl. Biased results of the direction and magnitude observed may lead to inappropriate physician action. Although there were no reports of affected patient sample results, it cannot be confirmed that patient sample results were not affected and would not be affected if the event were to recur undetected. There were no allegations of patient harm. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319809-2015-00125
MDR Report Key5316455
Date Received2015-12-21
Date of Report2015-12-21
Date of Event2015-11-27
Date Mfgr Received2015-11-25
Device Manufacturer Date2014-07-27
Date Added to Maude2015-12-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street513 TECHNOLOGY BLVD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS CHEMISTY PRODUCTS AMON SLIDES
Generic NameIN-VITRO DIAGNOSTIC
Product CodeJID
Date Received2015-12-21
Catalog Number1726926
Lot Number1014-0237-5676
Device Expiration Date2016-01-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-21

© 2024 FDA.report