[34133025] A customer in (b)(6) reported different calibration test results after testing twice when using vidas testosterone ii ref: (b)(4) lot 1004052240 expiry date: 08/16/2016. Although ref (b)(4) is not sold within the united states, a similar product, ref (b)(4), is sold within the united states. There were no reports of patient harm. An investigation will be conducted into this incident.
Patient Sequence No: 1, Text Type: D, B5


[38313175] An investigation into a discrepant calibration result with the vidas? Testosterone ii kit was performed. The customer did not return affected kit; therefore it was unable to be tested. A retain sample from the same lot was tested. This testing resulted in successful calibration results of s1= 1788 rfv and s1 = 1793 rfv. A manufacturing batch record review was performed and revealed no associated incidents or anomalies. Additionally, a complaint analysis for this lot was performed and revealed no other reported complaints. Based on the results of the investigation, the customer's complaint cannot be confirmed. The vidas? Testosterone ii kit is performing as intended.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2015-00173
MDR Report Key5318618
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2015-12-22
Date of Report2015-11-29
Date Mfgr Received2015-11-29
Device Manufacturer Date2015-06-11
Date Added to Maude2015-12-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX, S.A.
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? TESTOSTERONE II
Generic NameVIDAS? TESTOSTERONE II
Product CodeCDZ
Date Received2015-12-22
Catalog Number414320
Lot Number1004052240
Device Expiration Date2016-08-16
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.