ARCHITECT C4000 ANALYZER 02P24-40

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-12-24 for ARCHITECT C4000 ANALYZER 02P24-40 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[34936717] (b)(4). A followup report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[34936718] The account stated 2 patient samples generated elevated architect co2 results when processing on the pm shift that were repeated on the am shift with lower values. Id (b)(6) generated elevated architect co2 of 31. 5 meq/l but repeated 23. 3 meq/l. Id (b)(6) generated elevated architect co2 of 31. 9 meq/l but repeated 23. 5 meq/l. The account uses a reference range of 22 to 29 meq/l (19 to 59 years old) and 23 to 31 meq/l (60 years old and above). No specific patient information was provided. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


[36866879] Additional specific patient information was provided. Sample id (b)(6) is a female with date of birth 1986. Sample id (b)(6) is a female with date of birth 1974. Evaluation in process.
Patient Sequence No: 1, Text Type: N, H10


[38096192] After further evaluation, the suspect medical device was changed from architect c4000, list 02p24-40, manufacturing site (b)(6), to carbon dioxide, list 03l80-21, manufacturing site (b)(6). Mdr number 1415939-2016-00024 has been submitted and all further information will be documented under that mdr number.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1628664-2015-00337
MDR Report Key5326480
Date Received2015-12-24
Date of Report2016-02-11
Date of Event2015-12-02
Date Mfgr Received2016-01-15
Date Added to Maude2015-12-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT C4000 ANALYZER
Generic NameAUTOMATED CHEMISTRY ANALYZER
Product CodeCHS
Date Received2015-12-24
Catalog Number02P24-40
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.