LEICA BIOSYSTEMS PB0829

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-12-23 for LEICA BIOSYSTEMS PB0829 manufactured by Leica Biosystems Newcastle Ltd.

Event Text Entries

[34934467] The customer informed leica biosystems newcastle ltd that the tissue was stained again with a different unit of pb0829 from a different lot and demonstrated the correct staining. A diagnosis was made without having to perform a re biopsy, but there was a delay due to retesting the patient tissue. This product is sold as an analyte specific reagent in the us.
Patient Sequence No: 1, Text Type: N, H10


[34934468] Leica biosystems received a complaint regarding a (b)(6) high risk probe (product code pb0829) not staining on a known positive control. The complaint stated 5 different positive controls had been used and demonstrated negative staining.
Patient Sequence No: 1, Text Type: D, B5


[38883782] The retained unit has been tested at the (b)(4) facility using a known hpv positive tissue and appropriate reagent controls. No issues were observed with the resulting sections. The negative and positive controls were as expected. No further action required. The customer has been contacted several times to send back their unit but the customer has decided not to return the unit.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004859032-2015-00005
MDR Report Key5332286
Date Received2015-12-23
Date of Report2015-11-27
Date of Event2015-11-27
Date Mfgr Received2015-11-27
Device Manufacturer Date2014-11-19
Date Added to Maude2015-12-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNATALIE HENDERSON
Manufacturer StreetBALLIOL BUSINESS PARK WEST BENTON LANE
Manufacturer CityNEWCASTLE NE128EW
Manufacturer CountryUK
Manufacturer PostalNE12 8EW
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEICA BIOSYSTEMS
Generic NameBOND READ-TO-USE ISH HPV PROBE
Product CodeMVU
Date Received2015-12-23
Catalog NumberPB0829
Lot Number29078
Device Expiration Date2016-02-16
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS NEWCASTLE LTD
Manufacturer AddressBALLIOL BUSINESS PARK WEST BENTON LANE NEWCASTLE NE128EW UK NE12 8EW


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2015-12-23