STATIC AND DYNAMIC COMPLIANCE 22D E 777405-101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-12-29 for STATIC AND DYNAMIC COMPLIANCE 22D E 777405-101 manufactured by Carefusion.

Event Text Entries

[35063588] (b)(4). The customer replaced the screw of the table mount arm bracket with a longer screw which resolved the reported issue. The old screw was thought to be thrown away and will not be returning to carefusion for an evaluation. The device is currently working without issue. If more information becomes available a follow up will be submitted. Carefusion continues to track and trend any incident related to this issue. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[35063589] The customer reported the arm of the table mount became loose because the screws were too short. The arm never fell off because the staff tightened the screws 2-3 times per week. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2015-02563
MDR Report Key5333507
Date Received2015-12-29
Date of Report2015-12-04
Date of Event2015-12-04
Date Mfgr Received2015-12-04
Device Manufacturer Date2005-06-27
Date Added to Maude2015-12-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRESPIRATORY THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJILL RITTORNO
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer Phone8473628056
Manufacturer G1CAREFUSION
Manufacturer Street1100 BIRD CENTER DRIVE
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2015-12-29
Model Number22D E
Catalog Number777405-101
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2015-12-29

© 2026 FDA.report