ENDOSCOPE, NEPHROSCOPE, 25? 8972.433

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2015-12-29 for ENDOSCOPE, NEPHROSCOPE, 25? 8972.433 manufactured by Richard Wolf Gmbh.

Event Text Entries

[34647847] The medical device shows signs of mechanical damage. The working channel was deformed. The optic was produced in june 2011 and delivered to richard wolf (b)(4) in july 2011. The resolve of the adhesive bond is caused at wear and tear due to numerous treatment cycles. The user is informed about a visual and functional check in chapter 7 in the ifu ("run through the checks before and after each use. "). By doing these checks deficiencies or damages can be identified to protect the user from applying this instrument. To our knowledge here is no explainable relationship between the loss of the lens and the removal of the kidney.
Patient Sequence No: 1, Text Type: N, H10


[34647848] Richard wolf gmbh was informed in writing from richard wolf (b)(4) about the incident. The distal lens of the 8972,433 fell off during an intervention with ems lithoclast. The lens ended in the kidney of the patient. The kidney was removed completely.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611102-2015-00012
MDR Report Key5333724
Report SourceDISTRIBUTOR
Date Received2015-12-29
Date of Report2015-11-27
Date Mfgr Received2015-11-27
Device Manufacturer Date2011-06-09
Date Added to Maude2015-12-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. OLIVER EHRLICH
Manufacturer Street32 PFORZHEIMER STREET
Manufacturer CityKNITTLINGEN, 75438
Manufacturer CountryGM
Manufacturer Postal75438
Manufacturer Phone043351013
Manufacturer G1RICHARD WOLF GMBH
Manufacturer Street32 PFORZHEIMER STREET
Manufacturer CityKNITTLINGEN, 75438
Manufacturer CountryGM
Manufacturer Postal Code75438
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENDOSCOPE, NEPHROSCOPE, 25?
Generic NameNEPHROSCOPE
Product CodeGCM
Date Received2015-12-29
Model Number8972.433
OperatorPHYSICIAN
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRICHARD WOLF GMBH
Manufacturer Address32 PFORZHEIMER STREET KNITTLINGEN, 75438 GM 75438


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Other 2015-12-29