COMBO TRAY ADJUSTABLE TIBIAL TAMP HEAD, NET SHAPE 213-75-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-01-06 for COMBO TRAY ADJUSTABLE TIBIAL TAMP HEAD, NET SHAPE 213-75-01 manufactured by Exactech, Inc.

Event Text Entries

[35159143] The engineering evaluation noted that the experience reported was likely the result of excessive force un impaction, which lead to a brittle fracture of the locking feature. Device not returned.
Patient Sequence No: 1, Text Type: N, H10


[35159144] The impact head broke at the modular portion. The keel head was unable to extracted from the proximal tibial bone. This event occurred outside of the us.
Patient Sequence No: 1, Text Type: D, B5


[73478180] The engineering evaluation noted that the experience reported was likely the result of excessive force on impaction, which lead to a brittle fracture of the locking feature. Exactech confirmed that no pieces of the device were left in the patient.
Patient Sequence No: 1, Text Type: N, H10


[73478181] The impact head broke at the modular portion. The keel head was unable to extracted from the proximal tibial bone. This event occurred outside of the us.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1038671-2016-00023
MDR Report Key5347305
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-01-06
Date of Report2017-04-04
Date of Event2015-02-04
Date Mfgr Received2015-02-04
Date Added to Maude2016-01-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR GRAHAM CUTHBERT
Manufacturer Street2320 NW 66TH CT
Manufacturer CityGAINESVILLE FL 32653
Manufacturer CountryUS
Manufacturer Postal32653
Manufacturer Phone3523771140
Manufacturer G1EXACTECH, INC
Manufacturer Street2320 NW 66TH CT
Manufacturer CityGAINESVILLE FL 32653
Manufacturer CountryUS
Manufacturer Postal Code32653
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOMBO TRAY ADJUSTABLE TIBIAL TAMP HEAD, NET SHAPE
Generic NameTIBIAL TAMP HEAD
Product CodeHXG
Date Received2016-01-06
Catalog Number213-75-01
Lot Number61872001
OperatorPHYSICIAN
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerEXACTECH, INC
Manufacturer Address2320 NW 66TH CT GAINESVILLE FL 32653 US 32653


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-01-06

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.