TRINKLE QUICK-COUPLING MODIFIED FOR SMALL BATTERY DRIVE 532.020

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-01-06 for TRINKLE QUICK-COUPLING MODIFIED FOR SMALL BATTERY DRIVE 532.020 manufactured by Depuy Synthes Power Tools.

Event Text Entries

[35609209] Additional narrative: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment. Reliability engineering evaluated the device and the reported condition was confirmed. However, the assignable root cause was not determined. If additional information should become available, a supplemental medwatch report will be sent accordingly.
Patient Sequence No: 1, Text Type: N, H10


[35609210] It was reported by (b)(6) that during service and evaluation, it was observed that the coupling power supply on the quick coupling device was not functioning and defective. It was further determined that the attachment broke during shifting at motor clutch. Ԉis event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention or prolonged hospitalization. The exact date of this event was unknown. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8030965-2015-12813
MDR Report Key5347598
Date Received2016-01-06
Date of Report2015-12-08
Date Mfgr Received2015-12-22
Device Manufacturer Date2002-01-31
Date Added to Maude2016-01-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationSERVICE AND TESTING PERSONNEL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLINDA PLEWS
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone6107195000
Manufacturer G1DEPUY SYNTHES POWER TOOLS
Manufacturer StreetHAUPTSTRASSE 24
Manufacturer CityWALDENBURG CH4437
Manufacturer CountrySZ
Manufacturer Postal CodeCH4437
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTRINKLE QUICK-COUPLING MODIFIED FOR SMALL BATTERY DRIVE
Generic NameDRIVER, SURGICAL, PIN
Product CodeGFC
Date Received2016-01-06
Returned To Mfg2015-12-08
Catalog Number532.020
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES POWER TOOLS
Manufacturer AddressHAUPTSTRASSE 24 WALDENBURG CH4437 SZ CH4437


Patients

Patient NumberTreatmentOutcomeDate
10 2016-01-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.