CHROMID(TM) CPS? ELITE AGAR 418284

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-01-06 for CHROMID(TM) CPS? ELITE AGAR 418284 manufactured by Biomerieux Sa.

Event Text Entries

[35215711] A customer in (b)(6) reported discrepant results when using chromid(tm) cps elite agar; presence of a pink colony on cpse from urinary sample. There was no patient harm reported, however; there was a delay in the results due to the complementary testing.
Patient Sequence No: 1, Text Type: D, B5


[38799314] An investigation into a misidentification when using the chromid cps? Elite agar was performed by biomerieux. The organism in question had an enzymatic profile performed on it. The enzymatic profile revealed that this particular enterobacter organism expressed ss-galactosidase. Based on the design of the chromid cps? Elite agar any organism that produces ss-galactosidase will produce a red to burgundy coloration of the medium. This red to burgundy coloration is most typical of e. Coli organisms and an enterobacter organism does not typically produce ss-galactosidase. Based on the enzymatic profile of the organism, the product is performing as expected. The package insert states that certain species may produce colonies of the same characteristic color as e. Coli (ex. : citrobacter) as a product limitation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2016-00002
MDR Report Key5349020
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-01-06
Date of Report2015-12-16
Date Mfgr Received2015-12-16
Device Manufacturer Date2015-10-26
Date Added to Maude2016-01-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID(TM) CPS? ELITE AGAR
Generic NameCHROMID(TM) CPS? ELITE AGAR
Product CodeJXA
Date Received2016-01-06
Catalog Number418284
Lot Number1004399230
Device Expiration Date2016-02-21
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2016-01-06

© 2026 FDA.report