APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR 4407205

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2016-01-11 for APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR 4407205 manufactured by Life Technologies Holdings Pte Ltd..

Event Text Entries

[35623005] On (b)(4) 2015 a field service engineer visited the customer site to address the customer complaint of wavy data being reported from the instrument. The field service engineer inspected the instrument and replaced the existing lamp cable with a thicker cable. After the lamp cable was replaced, the engineer ran performance checks on the instrument. The instrument passed performance check specifications and the customer signed an acknowledgement of the service and repair.
Patient Sequence No: 1, Text Type: N, H10


[35623006] On (b)(6) 2015, a customer submitted a complaint regarding a 7500 fast dx instrument reporting "wavy data. " (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[38100971] An internal investigation of the complaint found that the issue of "wavy data" caused the affected assay run to fail background calibration. This would cause the run to be completed with no results reported. A historical review of the instrument was conducted. The instrument was installed on june 25, 2012. The replacement of the lamp cable on (b)(6) 2015 was the first instance the lamp cable was replaced. The cause of the replacement was attributed to normal wear and tear of the part. No further investigation is being conducted on the instrument.
Patient Sequence No: 1, Text Type: N, H10


[38100972] On (b)(6) 2015, a customer submitted a complaint regarding a 7500 fast dx instrument reporting "wavy data. " internal reference# (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003673482-2016-00001
MDR Report Key5359368
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2016-01-11
Date of Report2015-12-14
Date of Event2015-12-14
Date Mfgr Received2015-12-14
Date Added to Maude2016-01-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KELLI TANZELLA
Manufacturer Street3175 STALEY ROAD
Manufacturer CityGRAND ISLAND NY 14072
Manufacturer CountryUS
Manufacturer Postal14072
Manufacturer Phone7167743122
Manufacturer G1LIFE TECHNOLOGIES HOLDINGS PTE LTD.
Manufacturer StreetBLK 33, MARSILING INDUSTRIAL ESTATE ROAD 3, NO. 07-06
Manufacturer CitySINGAPORE NORTH WEST, SINGAPORE 739256
Manufacturer CountrySN
Manufacturer Postal Code739256
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAPPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR
Generic NameAPPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR
Product CodeNSU
Date Received2016-01-11
Catalog Number4407205
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES HOLDINGS PTE LTD.
Manufacturer AddressBLK 33, MARSILING INDUSTRIAL ESTATE ROAD 3, NO. 07-06 SINGAPORE NORTH WEST, SINGAPORE 739256 SN 739256


Patients

Patient NumberTreatmentOutcomeDate
10 2016-01-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.