MGT MOVEMENT PROX. PHAL. TOTAL SZ. 40 MGT-890-40PPT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2016-01-13 for MGT MOVEMENT PROX. PHAL. TOTAL SZ. 40 MGT-890-40PPT manufactured by Ascension Orthopedics.

Event Text Entries

[35757069] It was reported the implant backed out for some unknown reason and the patient required a revision. It was reported the event is surgeon error. "there was nothing wrong with the implant. It just backed out and the joint dislocated. "
Patient Sequence No: 1, Text Type: D, B5


[40407271] Additional information received 8mar16: the product was used. It won't be returned. There was nothing wrong with it. Integra completed its internal investigation 25feb16. The investigation included: method: review of device history records. Review of complaint management database for similar complaints. Results: no product/manufacturing issues were identified that would have caused or contributed to the event. A review of the complaint files found one complaint for an mgt proximal phalanx backing out. (b)(4). Conclusion: as the implant was not returned, a definitive root cause cannot be established. Possible root causes may include not following the surgical technique, or other surgeon errors.
Patient Sequence No: 1, Text Type: N, H10


[40407272] Corrected event type to adverse event. Information provided indicates there was no product problem.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1651501-2016-00001
MDR Report Key5362982
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2016-01-13
Date of Report2015-12-22
Date of Event2015-12-15
Date Mfgr Received2016-02-25
Date Added to Maude2016-01-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER MARIA LEONARD
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1ASCENSION ORTHOPEDICS
Manufacturer Street8700 CAMERON ROAD #100
Manufacturer CityAUSTIN TX 78754
Manufacturer CountryUS
Manufacturer Postal Code78754
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMGT MOVEMENT PROX. PHAL. TOTAL SZ. 40
Generic NameMOVEMENT GREAT TOE
Product CodeKWD
Date Received2016-01-13
Catalog NumberMGT-890-40PPT
Lot Number15-2669
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerASCENSION ORTHOPEDICS
Manufacturer Address8700 CAMERON ROAD #100 8700 CAMERON ROAD #100 AUSTIN TX 78754 US 78754


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-01-13

© 2026 FDA.report