DUROGRIP NEEDLE HOLDER STR.D:5/310MM PL407R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2016-01-14 for DUROGRIP NEEDLE HOLDER STR.D:5/310MM PL407R manufactured by Aesculap Ag.

Event Text Entries

[36047049] (b)(4). Manufacturing site evaluation: on-going.
Patient Sequence No: 1, Text Type: N, H10


[36047050] Country of complaint: (b)(6). The metal part which goes into the handle is not closing. There was no patient injury but a 30 minute delay in surgery as there were no others available to use. Additional report submitted at the same time, same facility, same event description. See 2916714-2016-00022.
Patient Sequence No: 1, Text Type: D, B5


[53395051] The needle holders are in a used condition. The tips of the locking pin are broken off. The fragments are not available. The instruments were analyzed by microscope. No pores, inclusions or foreign bodies could be found. The device quality and manufacturing history records are not possible. Based on the information available as well as a result of the investigation, root cause of the failure is most probably a material error. The current failure rate is within the risk analysis and therefore acceptable.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2016-00023
MDR Report Key5369009
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-01-14
Date of Report2016-08-08
Date Facility Aware2016-01-11
Date Mfgr Received2016-06-23
Device Manufacturer Date2015-03-01
Date Added to Maude2016-01-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1AESCULAP AG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN, 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDUROGRIP NEEDLE HOLDER STR.D:5/310MM
Generic NameNEEDLE HOLDER
Product CodeHXK
Date Received2016-01-14
Returned To Mfg2016-01-14
Model NumberPL407R
Catalog NumberPL407R
Device AvailabilityR
Device Age17 MO
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG
Manufacturer AddressPO BOX 40 TUTTLINGEN, 78501 GM 78501


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-01-14

© 2026 FDA.report