COBAS TAQSCREEN MPX TEST V2.0 CE-IVD 05969492190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-01-15 for COBAS TAQSCREEN MPX TEST V2.0 CE-IVD 05969492190 manufactured by Roche Molecular Systems.

Event Text Entries

[36483968] The investigation into this issue is currently on-going. The outcome of this investigation will be communicated through a follow-up report. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[36483971] A blood bank in (b)(6) filed a complaint stating that (b)(6) results for one blood screening donor sample were generated when using the cobas taqscreen mpx v2. 0 ce-ivd test. The initial pool of one results were (b)(6). The repeat pool gave (b)(6). The final result was reported as (b)(6). All testing was performed on (b)(6) 2015. The serology results for this donor are currently unknown.
Patient Sequence No: 1, Text Type: D, B5


[37747494] The customer from (b)(6) alleged discrepant hbv results for a donor when tested with the taqscreen mpx test, v2. 0. The patient's serology results were later found to be negative. The investigation determined that the complaint kit lot met release specifications. Additionally, the instrumentation and algorithm used performed as intended. The initial reactive result was most likely due to sample specific issues.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2243471-2016-00008
MDR Report Key5370218
Date Received2016-01-15
Date of Report2016-01-29
Date of Event2015-12-03
Date Mfgr Received2015-12-17
Date Added to Maude2016-01-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA VINCENT STAGNITTO
Manufacturer Street1080 US HWY 202 S NA
Manufacturer CityBRANCHBURG NJ 08876
Manufacturer CountryUS
Manufacturer Postal08876
Manufacturer Phone9082537569
Manufacturer G1NA
Manufacturer StreetNA NA
Manufacturer CityNA
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS TAQSCREEN MPX TEST V2.0 CE-IVD
Generic NameASSAY, HYB AND/OR NUCLEIC ACID AMP FOR DETECTION OF HCV RNA, HIV RNA
Product CodeMZP
Date Received2016-01-15
Model NumberNA
Catalog Number05969492190
Lot Number185527
ID NumberNA
Device Expiration Date2016-12-31
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE MOLECULAR SYSTEMS
Manufacturer Address1080 US HWY 202 S NA BRANCHBURG NJ 088763733 US 088763733


Patients

Patient NumberTreatmentOutcomeDate
10 2016-01-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.