PALL BDS SAMPLE SET BDS 02 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2004-03-23 for PALL BDS SAMPLE SET BDS 02 NA manufactured by Pall Biomedical Products Co..

Event Text Entries

[338139] It was reported that during use in a non-clinical setting, one(1) of two(2) devices undergoing process validation testing at a user facility gave unexpected readings which led to an erroneous conclusion about the sample being tested (false negative for bacterial contamination), during qc evaluation of bb processes. When the two(2) instruments that were used for validation testing again the next day, after replacement of the disposable sampling tube assemblies, both instruments gave comparable expected readings.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2432733-2004-00001
MDR Report Key537673
Report Source05,06,07
Date Received2004-03-23
Date of Report2004-02-24
Date of Event2004-02-20
Date Facility Aware2004-02-20
Date Reported to Mfgr2004-02-24
Date Mfgr Received2004-02-24
Date Added to Maude2004-08-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLEONARD BERMAN, PH.D
Manufacturer Street2200 NORTHERN BOULEVARD
Manufacturer CityEAST HILLS NY 11548
Manufacturer CountryUS
Manufacturer Postal11548
Manufacturer Phone*
Manufacturer G1PALL BIOMEDICAL PRODUCTS COMPANY
Manufacturer Street2200 NORTHERN BOULEVARD
Manufacturer CityEAST HILLS NY 11548
Manufacturer CountryUS
Manufacturer Postal Code11548
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePALL BDS SAMPLE SET
Generic NameBACTERIA DETECTION SYSTEM
Product CodeMZC
Date Received2004-03-23
Model NumberBDS 02
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age1 YR
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key526924
ManufacturerPALL BIOMEDICAL PRODUCTS CO.
Manufacturer Address2200 NORTHERN BLVD. EAST HILLS NY 11548 US
Baseline Brand NamePALL BDS SAMPLE SET
Baseline Generic NameBACTERIA DETECTION SYSTEM
Baseline Model NoBDS 02
Baseline Catalog NoNA
Baseline IDNA
Baseline Device FamilyBACTERIA DETECTION SYSTEM
Baseline Shelf Life [Months]NA
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK020014
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN