SODASORB CARBON DIOXIDE ABSORBENT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-01-18 for SODASORB CARBON DIOXIDE ABSORBENT manufactured by Grace Discovery Sciences.

Event Text Entries

[36883406] While prepping the pt, anesthesiologist expressed concerns about increasing etco2 levels, possible malignant hyperthermia. Malignant hyperthermia cart immediately brought in. Etco2 went from 30% to 50%. No temperature elevation noted. Foley cath draining yellow urine. Pt given 8 vials of dantrolene. Istat labs drawn sicu called report, cooling blanket. Per intensivist: except for hypercarbia, pt has no tachycardia, no muscle rigidity and no hyperthermia, and the cpk levels were normal. Suspicion for malignant hyperthermia is very less at this point. Believed upon further investigation, this may have been caused by the sodasorb filter turning to violet, then back to white without color change being noted.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5059548
MDR Report Key5384780
Date Received2016-01-18
Date of Report2016-01-18
Date of Event2015-11-12
Date Added to Maude2016-01-22
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSODASORB CARBON DIOXIDE ABSORBENT
Generic NameSODASORB CARBON DIOXIDE ABSORBENT
Product CodeCBL
Date Received2016-01-18
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerGRACE DISCOVERY SCIENCES
Manufacturer AddressCOLUMBIA MD 21044 US 21044


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2016-01-18