MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-01-22 for I-STAT DOWNLOADER RECHARGER 04P73-04 manufactured by Abbott Point Of Care.
[36754043]
Apoc incident # (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[36754044]
On (b)(6) 2016, abbott point of care (apoc) was contacted by a customer regarding an analyzer that was placed in the downloader/recharger (drc) and a "pop" was heard. The operator states that smoke was observed along with the burning smell. The customer states that the downloader caused the event and opted to return downloader and cables for investigation. The product was replaced at no charge. Per i-stat system manual: art: 714368-00k, rev. Date: 02-aug-2012: the downloader/recharger can recharge a rechargeable battery in the analyzer. If the analyzer contains a rechargeable battery, the battery begins recharging automatically as soon as the analyzer is placed in the downloader/recharger. The downloader/recharger also has a compartment for recharging a rechargeable battery outside the analyzer. Placing an analyzer in a downloader/recharger will automatically initiate recharging of the rechargeable battery. The indicator light on top of the downloader/recharger will be green (trickle charge), red (fast charge), or blinking red (fast charge pending) when an analyzer with a rechargeable battery is placed in the downloader/recharger. As well, placing a rechargeable battery into the recharging compartment will automatically initiate trickle recharging. The indicator light near the recharging compartment will be green when a rechargeable battery is placed in the compartment. There are no injuries associated with this event. At this time there is no evidence of a confirmed malfunction and product was replaced at no charge.
Patient Sequence No: 1, Text Type: D, B5
[42409409]
Apoc incident # (b)(4). The investigation was completed on 03/31/2016. The cause of the customer complaint was due to an electronic component failure.
Patient Sequence No: 1, Text Type: N, H10
[42409410]
Na.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2245578-2016-00001 |
MDR Report Key | 5386325 |
Date Received | 2016-01-22 |
Date of Report | 2016-04-11 |
Date of Event | 2016-01-05 |
Date Mfgr Received | 2016-03-31 |
Device Manufacturer Date | 2015-10-18 |
Date Added to Maude | 2016-01-22 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | LINDA MACZUSZENKO |
Manufacturer Street | 400 COLLEGE ROAD |
Manufacturer City | PRINCETON NJ 08540 |
Manufacturer Country | US |
Manufacturer Postal | 08540 |
Manufacturer Phone | 6136885949 |
Manufacturer G1 | FLEXTRONICS MANUFACTURING (SINGAPORE) |
Manufacturer Street | PTE LTD 1 KALLANG PLACE |
Manufacturer City | SINGAPORE 339211 |
Manufacturer Country | SN |
Manufacturer Postal Code | 339211 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | I-STAT DOWNLOADER RECHARGER |
Generic Name | DOWNLOADER/RECHARGER |
Product Code | JBP |
Date Received | 2016-01-22 |
Returned To Mfg | 2016-01-22 |
Model Number | NA |
Catalog Number | 04P73-04 |
Lot Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT POINT OF CARE |
Manufacturer Address | 400 COLLEGE ROAD PRINCETON NJ 085406607 US 085406607 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2016-01-22 |