I-STAT DOWNLOADER RECHARGER 04P73-04

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-01-22 for I-STAT DOWNLOADER RECHARGER 04P73-04 manufactured by Abbott Point Of Care.

Event Text Entries

[36754043] Apoc incident # (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[36754044] On (b)(6) 2016, abbott point of care (apoc) was contacted by a customer regarding an analyzer that was placed in the downloader/recharger (drc) and a "pop" was heard. The operator states that smoke was observed along with the burning smell. The customer states that the downloader caused the event and opted to return downloader and cables for investigation. The product was replaced at no charge. Per i-stat system manual: art: 714368-00k, rev. Date: 02-aug-2012: the downloader/recharger can recharge a rechargeable battery in the analyzer. If the analyzer contains a rechargeable battery, the battery begins recharging automatically as soon as the analyzer is placed in the downloader/recharger. The downloader/recharger also has a compartment for recharging a rechargeable battery outside the analyzer. Placing an analyzer in a downloader/recharger will automatically initiate recharging of the rechargeable battery. The indicator light on top of the downloader/recharger will be green (trickle charge), red (fast charge), or blinking red (fast charge pending) when an analyzer with a rechargeable battery is placed in the downloader/recharger. As well, placing a rechargeable battery into the recharging compartment will automatically initiate trickle recharging. The indicator light near the recharging compartment will be green when a rechargeable battery is placed in the compartment. There are no injuries associated with this event. At this time there is no evidence of a confirmed malfunction and product was replaced at no charge.
Patient Sequence No: 1, Text Type: D, B5


[42409409] Apoc incident # (b)(4). The investigation was completed on 03/31/2016. The cause of the customer complaint was due to an electronic component failure.
Patient Sequence No: 1, Text Type: N, H10


[42409410] Na.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2245578-2016-00001
MDR Report Key5386325
Date Received2016-01-22
Date of Report2016-04-11
Date of Event2016-01-05
Date Mfgr Received2016-03-31
Device Manufacturer Date2015-10-18
Date Added to Maude2016-01-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLINDA MACZUSZENKO
Manufacturer Street400 COLLEGE ROAD
Manufacturer CityPRINCETON NJ 08540
Manufacturer CountryUS
Manufacturer Postal08540
Manufacturer Phone6136885949
Manufacturer G1FLEXTRONICS MANUFACTURING (SINGAPORE)
Manufacturer StreetPTE LTD 1 KALLANG PLACE
Manufacturer CitySINGAPORE 339211
Manufacturer CountrySN
Manufacturer Postal Code339211
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameI-STAT DOWNLOADER RECHARGER
Generic NameDOWNLOADER/RECHARGER
Product CodeJBP
Date Received2016-01-22
Returned To Mfg2016-01-22
Model NumberNA
Catalog Number04P73-04
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT POINT OF CARE
Manufacturer Address400 COLLEGE ROAD PRINCETON NJ 085406607 US 085406607


Patients

Patient NumberTreatmentOutcomeDate
10 2016-01-22