FINESSE PATCH

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,study report with the FDA on 2016-01-25 for FINESSE PATCH manufactured by Calibra Medical, Inc.

Event Text Entries

[36779167] Calibra has been unable to request return of the product at this time. If the device is returned, an evaluation shall be completed and a supplemental report will be filed. No conclusions can be made at this time. (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[36779168] On (b)(6) 2016 it was reported that the patient's blood glucose was 184mg/dl with no signs or symptoms of hyperglycemia after taking a dose of insulin from the device. There was no medical intervention; the patient replaced the device with a new one, dosed with insulin, and hyperglycemia resolved. This hyperglycemic incident does not meet the definition of an adverse event. It was said that when the patient removed the device from the skin, the cannula was bent. This complaint is being reported because the device was unable to deliver insulin due to a bent cannula.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008272700-2016-00006
MDR Report Key5388877
Report SourceCONSUMER,STUDY
Date Received2016-01-25
Date of Report2016-01-20
Date Mfgr Received2016-01-20
Date Added to Maude2016-01-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSAM CRAWFORD
Manufacturer Street220 SAGINAW DR
Manufacturer CityREDWOOD CITY CA 940634725
Manufacturer CountryUS
Manufacturer Postal940634725
Manufacturer Phone6502984705
Manufacturer G1CALIBRA MEDICAL, INC
Manufacturer Street220 SAGINAW DR
Manufacturer CityREDWOOD CITY CA 940634725
Manufacturer CountryUS
Manufacturer Postal Code940634725
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFINESSE PATCH
Generic NameFINESSE INSULIN DELIVERYSYSTEM
Product CodeOPP
Date Received2016-01-25
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerCALIBRA MEDICAL, INC
Manufacturer Address220 SAGINAW DR REDWOOD CITY CA 940634725 US 940634725


Patients

Patient NumberTreatmentOutcomeDate
10 2016-01-25