LIGASURE BLUNT TIP LAPAROSCOPIC SEALER/DIVIDER LF1637

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-01-25 for LIGASURE BLUNT TIP LAPAROSCOPIC SEALER/DIVIDER LF1637 manufactured by Covidien Lp.

Event Text Entries

[36928890] (b)(4). The incident device has been received and is under evaluation. When the device evaluation is complete a follow-up report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[36928891] The customer reported that during a laparoscopic gyn case the ligasure did not seal or burn. The device was recognized by the generator and "appropriate sounds" were heard during activation. There was no injury to the patient.
Patient Sequence No: 1, Text Type: D, B5


[39029392] (b)(4). Date of follow-up report : 02/25/2016. The incident device was returned, evaluated and found to function within specification. The device was tested for activation and sealing function on simulated tissue with acceptable results. Additional testing was done by performing multiple seals of various sizes on porcine tissue and all seal cycles were completed satisfactorily. Covidien was unable to confirm the customer? S report.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1717344-2016-00063
MDR Report Key5389605
Date Received2016-01-25
Date of Report2015-12-30
Date of Event2015-12-28
Date Mfgr Received2016-02-05
Date Added to Maude2016-01-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS SHARON MURPHY
Manufacturer Street5920 LONGBOW DRIVE
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal80301
Manufacturer Phone2034925267
Manufacturer G1COVIDIEN LP
Manufacturer Street5920 LONGBOW DRIVE
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal Code80301
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLIGASURE BLUNT TIP LAPAROSCOPIC SEALER/DIVIDER
Generic NameLIGASURE VESSEL SEALING SYSTEM
Product CodeBWA
Date Received2016-01-25
Returned To Mfg2016-01-15
Model NumberLF1637
Catalog NumberLF1637
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOVIDIEN LP
Manufacturer Address5920 LONGBOW DRIVE BOULDER CO 80301 US 80301


Patients

Patient NumberTreatmentOutcomeDate
10 2016-01-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.