SCALP CLIP CARTRIDGE UNKNOWN FF013P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-01-28 for SCALP CLIP CARTRIDGE UNKNOWN FF013P manufactured by Aesculap, Inc..

Event Text Entries

[36986366]
Patient Sequence No: 1, Text Type: N, H10


[36986367] The scalp gun would not fire the scalp clips to attach to the patient's scalp. A new gun was opened and it still wouldn't work. The clip cartridge was replaced and it still wouldn't work properly. Two scalp clips fired but did not attach to the patient as appropriate. All four items were removed from the field with the exception of the 2 failed scalp clips which were visible in the craniotomy drape. Raney clip appliers were opened and raney clips were opened and used instead. Upon inspection, it appears that one gun was working properly but the cartridges would not fire on either gun. All items will be turned in for further inspection at the end of the case.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number5396021
MDR Report Key5396021
Date Received2016-01-28
Date of Report2015-11-18
Date of Event2015-01-23
Report Date2015-11-18
Date Reported to FDA2015-11-18
Date Reported to Mfgr2015-11-18
Date Added to Maude2016-01-28
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSCALP CLIP CARTRIDGE
Generic NameCLIP, SCALP
Product CodeHBO
Date Received2016-01-28
Returned To Mfg2015-04-28
Model NumberUNKNOWN
Catalog NumberFF013P
Lot NumberUNKNOWN
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP, INC.
Manufacturer Address3773 CORPORATE PARKWAY CENTER VALLEY PA 18034 US 18034


Patients

Patient NumberTreatmentOutcomeDate
10 2016-01-28

© 2026 FDA.report