Event Text Entries

[37729953] The investigation is on-going; the outcome of this investigation will be communicated through a follow-up report. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[37729954] A blood center in the us filed a complaint alleging generation of a (b)(6) for one donor while using the cobas taqscreen mpx test v2, lot w07772. The donor tested (b)(6) with the cobas ampliscreen test. The donor was (b)(6).
Patient Sequence No: 1, Text Type: D, B5


[44734014] A us customer reported the generation of a (b)(6) result for a donor sample when tested with the cobas taqscreen mpx test, v2. The donor generated positive serology results and a reactive result with another pcr test. Review of the pcr growth curve for the sample in question indicated that it was anomalous in nature. This type of growth curve has not been reported nor identified previously. Sufficient sample volume is not available for further investigation. Without the sample, the cause for the anomalous growth curve cannot be determined, and the sequence of the sample cannot be obtained and evaluated against the cobas taqscreen mpx test, v2 performance claims. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2243471-2016-00009
MDR Report Key5396450
Date Received2016-01-28
Date of Report2016-05-09
Date of Event2015-12-17
Date Mfgr Received2015-12-30
Date Added to Maude2016-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA VINCENT STAGNITTO
Manufacturer Street1080 US HWY 202 S NA
Manufacturer CityBRANCHBURG NJ 088763733
Manufacturer CountryUS
Manufacturer Postal088763733
Manufacturer Phone9082537569
Manufacturer G1NA
Manufacturer StreetNA NA
Manufacturer CityNA
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS TAQSCREEN MPX TEST, V2 US-IVD
Generic NameASSAY, HYB AND/OR NUCLEIC ACID AMP FOR DETECTION OF HCV RNA, HIV RNA
Product CodeMZP
Date Received2016-01-28
Model NumberNA
Catalog Number05969484190
Lot NumberW07772
ID NumberNA
Device Expiration Date2017-02-28
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE MOLECULAR SYSTEMS, INC
Manufacturer Address1080 US HIGHWAY 202 SOUTH NA BRANCHBURG NJ 088763733 US 088763733