RUSCH TRULED ADULT RECHARGEABLE BATTERY 0055502

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2016-01-28 for RUSCH TRULED ADULT RECHARGEABLE BATTERY 0055502 manufactured by Teleflex Medical.

Event Text Entries

[37024479] (b)(4). The device sample was not returned for evaluation at the time of this report.
Patient Sequence No: 1, Text Type: N, H10


[37024480] The customer alleges that the screw ring and battery metal tip has fallen off. The alleged issue was detected at the time of preparation for patient use. No patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


[42627233] (b)(4). The sample was not returned for evaluation; therefore, the complaint could not be confirmed. If the sample is returned, a follow-up report will be submitted with investigation results.
Patient Sequence No: 1, Text Type: N, H10


[42627234] The customer alleges that the screw ring and battery metal tip has fallen off. The alleged issue was detected at the time of preparation for patient use. No patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3011137372-2016-00007
MDR Report Key5396665
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2016-01-28
Date of Report2016-01-13
Date of Event2016-01-13
Date Mfgr Received2016-04-14
Date Added to Maude2016-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal Code27560
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRUSCH TRULED ADULT RECHARGEABLE BATTERY
Generic NameBATTERY, RECHARGEABLE
Product CodeFCO
Date Received2016-01-28
Catalog Number0055502
Lot NumberUNKNOWN
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2016-01-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.