HOTBAG

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-01-28 for HOTBAG manufactured by Check Corp..

Event Text Entries

[37426744] Hot bag blanket warmer started to smoke and melt on the back of the bag where the connectors are located in the bag. The power box melted. Staff were able to throw it outside before any flames or other damage. The heating element work order number was (b)(4). I notified the company and they are having their engineers review it and will send me a box to send the malfunctioning device back to them. I have 3 other hotbag blanket warmers that i inspected and feel that the connector insulation is brttle and i voluntarily pulled them from service. The company is aware of this also.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5059830
MDR Report Key5403668
Date Received2016-01-28
Date of Report2016-01-28
Date of Event2016-01-28
Date Added to Maude2016-02-02
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHOTBAG
Generic NameHOTBAG
Product CodeIRQ
Date Received2016-01-28
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCHECK CORP.
Manufacturer Address1800 STEPHENSON HWY. TROY MI 48083 US 48083


Patients

Patient NumberTreatmentOutcomeDate
10 2016-01-28

© 2025 FDA.report