HUDSON BITE-GARD MOLAR BITE BLOCK 1140

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-02-02 for HUDSON BITE-GARD MOLAR BITE BLOCK 1140 manufactured by Teleflex Medical.

Event Text Entries

[37338281] (b)(4). A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned. No photo available. A device history record review could not be conducted since the lot number was not provided. Customer complaint cannot be confirmed due to the lack of sample and no lot number provided for investigation. An attempt to duplicate the failure mode was made but at the time there is no inventory of the involved product code nor is it being manufactured at the time. If the device sample becomes available this investigation will be updated with the evaluation results.
Patient Sequence No: 1, Text Type: N, H10


[37338282] The customer alleges that the bite block broke off in the patient's mouth. The broken piece was retrieved from the patient's throat. No patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004365956-2016-00108
MDR Report Key5405110
Report SourceUSER FACILITY
Date Received2016-02-02
Date of Report2016-01-25
Date of Event2016-01-20
Date Mfgr Received2016-01-25
Date Added to Maude2016-02-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARGIE BURTON, RN
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334965
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetPARQUE INDUSTRIAL FINSA
Manufacturer CityNUEVO LAREDO 88275
Manufacturer CountryMX
Manufacturer Postal Code88275
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON BITE-GARD MOLAR BITE BLOCK
Generic NameENDO. BITE GARD AND ACCESSOR.
Product CodeMNK
Date Received2016-02-02
Catalog Number1140
Lot NumberUNKNOWN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-02-02