LIGASURE ADVANCE PISTOL GRIP LF5544

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-02-08 for LIGASURE ADVANCE PISTOL GRIP LF5544 manufactured by Covidien Lp.

Event Text Entries

[38239191] (b)(4). Date of initial report : 02/08/2016. Visual inspection found the knife is trapped and contained within the jaws. The jaws were stuck shut due to the knife is trapped and contained within the jaws. Knife trap happens when the blade is extended and the jaws are not completely closed. This allows the knife track to open too wide and the blade moves out of its track. As a safety measure, the knife must be retracted in order to open the jaws. The tissue in the webbing may have prevented the knife from retracting far enough to allow the jaws to open. More frequent cleaning could have also reduced the difficulty activating the knife. The ifu cautions confirm that the jaws have reached the closed position and are locked (the handle is latched) before activating the cutter. Manufacturing non-conformances were reviewed. No entries pertinent to the report were noted.
Patient Sequence No: 1, Text Type: N, H10


[38239192] The customer reported that the jaws of the device stopped opening wide during a laparoscopic colon procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1717344-2016-00115
MDR Report Key5417608
Date Received2016-02-08
Date of Report2016-01-12
Date of Event2016-01-11
Date Mfgr Received2016-01-12
Device Manufacturer Date2015-09-01
Date Added to Maude2016-02-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS SHARON MURPHY
Manufacturer Street5920 LONGBOW DRIVE
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal80301
Manufacturer Phone2034925267
Manufacturer G1COVIDIEN LP
Manufacturer Street5920 LONGBOW DRIVE
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal Code80301
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameLIGASURE ADVANCE PISTOL GRIP
Generic NameLIGASURE VESSEL SEALING SYSTEM
Product CodeBWA
Date Received2016-02-08
Returned To Mfg2016-01-21
Model NumberLF5544
Catalog NumberLF5544
Lot Number52180076X
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOVIDIEN LP
Manufacturer Address5920 LONGBOW DRIVE BOULDER CO 80301 US 80301


Patients

Patient NumberTreatmentOutcomeDate
10 2016-02-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.