OXYCON PRO 001-143960

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-02-10 for OXYCON PRO 001-143960 manufactured by Carefusion.

Event Text Entries

[38483155] (b)(4). This mdr is being submitted in response to a two year retrospective complaint review that was performed by carefusion to identify mdrs following the issuance of a 483. At this time the device is functioning as intended and is in normal operation. Should additional information become available a supplemental report will be submitted. Carefusion continues to track and trend any incident related to this issue. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[38483157] The customer reported that while doing a test the treadmill unexpectedly stopped. There was no patient harm or injury reported. A carefusion field service technician (fst) inspected the treadmill and all of its connections and ran a test protocol study without issue. All factors and settings were found to be within specifications. It was determined that the treadmill time out had to extended from 3000 to 10,000 milliseconds and a corrupted nonin 7600 driver was found and changed for a nonin 8600 driver.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2016-03156
MDR Report Key5428961
Date Received2016-02-10
Date of Report2014-02-12
Date of Event2014-02-12
Date Mfgr Received2014-02-12
Date Added to Maude2016-02-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRESPIRATORY THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJILL RITTORNO
Manufacturer Street22745 SAVI RANCH PKWY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone7149227830
Manufacturer G1CAREFUSION
Manufacturer Street1100 BIRD CENTER DRIVE
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameOXYCON PRO
Generic NameCALCULATOR, PULMONARY FUNCTION DATA
Product CodeBZC
Date Received2016-02-10
Model NumberOXYCON PRO
Catalog Number001-143960
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2016-02-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.