VIDAS? MEASLES IGG 30219

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-02-12 for VIDAS? MEASLES IGG 30219 manufactured by Biomerieux Sa.

Event Text Entries

[38142595] A customer at "(b)(6)" reported discrepant results in three patients with measles while using; vidas? Measles igg test.
Patient Sequence No: 1, Text Type: D, B5


[46266753] A customer in the united states reported discrepant results associated with vidas? Measles igg test lot 1003910510 and another method (quest). Results of the internal biom? Rieux investigation are as follows: the analysis of the batch history records showed no anomaly, but some positive sera were found outside of the internal norms; however, without changing of serological interpretations. No other complaint was recorded for vidas? Measles igg lot 1003910510/160213-0. No possibility to test samples due to the expiration of vidas? Measles igg lot 1003910510/160213-0 as of 13feb2016. Vidas? Measles igg performance does not claim 100% in terms of specificity or sensitivity. Without the quest performance data (sensitivity and specificity) and the lack of a third method, it is not possible to determine which result is correct between the two (2) methods. In conclusion, no product problem was identified for vidas? Measles igg lot 1003910510/160213-0.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2016-00033
MDR Report Key5434235
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-02-12
Date of Report2015-10-16
Date Mfgr Received2015-10-16
Device Manufacturer Date2015-04-13
Date Added to Maude2016-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? MEASLES IGG
Generic NameVIDAS? MEASLES IGG
Product CodeLJB
Date Received2016-02-12
Catalog Number30219
Lot Number1003910510
Device Expiration Date2016-02-13
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2016-02-12