MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-02-16 for ICONNECT ENTERPRISE ARCHIVE 9.1.2 manufactured by Merge Healthcare.
[38219246]
Upon retrospective review, this issue was determined to be reportable as an mdr.
Patient Sequence No: 1, Text Type: N, H10
[38219247]
Customer reported that priors are failing to retrieve back to facility image cache (fic) for one patient. The pacs admin retrieved the priors manually. User used qc function to make a change in the data and inadvertently connected the studies to another patient. Further analysis has found that at some point, the customer modified a group of studies on the long term archive (lta) at the site which resulted in the studies being placed under an incorrect patient. (b)(4).
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2183926-2015-00138 |
MDR Report Key | 5437880 |
Report Source | USER FACILITY |
Date Received | 2016-02-16 |
Date of Report | 2016-02-15 |
Date of Event | 2015-07-27 |
Date Mfgr Received | 2015-07-27 |
Device Manufacturer Date | 2014-01-10 |
Date Added to Maude | 2016-02-16 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. MICHAEL DIEDRICK |
Manufacturer Street | 900 WALNUT RIDGE RD |
Manufacturer City | HARTLAND WI 53029 |
Manufacturer Country | US |
Manufacturer Postal | 53029 |
Manufacturer Phone | 2629123570 |
Manufacturer G1 | MERGE HEALTHCARE |
Manufacturer Street | 900 WALNUT RIDGE DRIVE |
Manufacturer City | HARTLAND WI 53029 |
Manufacturer Country | US |
Manufacturer Postal Code | 53029 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ICONNECT ENTERPRISE ARCHIVE |
Generic Name | IMAGE ARCHIVE |
Product Code | LMB |
Date Received | 2016-02-16 |
Model Number | 9.1.2 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MERGE HEALTHCARE |
Manufacturer Address | 900 WALNUT RIDGE DRIVE HARTLAND WI 53029 US 53029 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2016-02-16 |