ICONNECT ENTERPRISE ARCHIVE 10.0.2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-02-16 for ICONNECT ENTERPRISE ARCHIVE 10.0.2 manufactured by Merge Healthcare.

Event Text Entries

[38227896] Upon retrospective review, this issue was determined to be reportable as an mdr. Development assessed and implemented a fix at the customer site. Customer reported no longer seeing the error message.
Patient Sequence No: 1, Text Type: N, H10


[38227897] Iconnect enterprise archive a vendor neutral archive for storage and communications of medical images and data. A customer was migrating from radsuite to merge pacs. During the migration, image load errors were occurring on studies after the half folder tool was ran. Customers migrating from radsuite to merge pacs will see this issue for any data stored prior to iconnect enterprise archive version 8. X. This issue causes image load errors in the pacs viewer preventing a user from viewing priors. The customer did not report a serious injury to the patient. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183926-2016-00002
MDR Report Key5437905
Report SourceUSER FACILITY
Date Received2016-02-16
Date of Report2016-02-15
Date of Event2015-11-09
Date Mfgr Received2015-11-09
Device Manufacturer Date2015-07-21
Date Added to Maude2016-02-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHAEL DIEDRICK
Manufacturer Street900 WALNUT RIDGE RD
Manufacturer CityHARTLAND WI 53029
Manufacturer CountryUS
Manufacturer Postal53029
Manufacturer Phone2629123570
Manufacturer G1MERGE HEALTHCARE
Manufacturer Street900 WALNUT RIDGE DRIVE
Manufacturer CityHARTLAND WI 53029
Manufacturer CountryUS
Manufacturer Postal Code53029
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameICONNECT ENTERPRISE ARCHIVE
Generic NameARCHIVER
Product CodeLMB
Date Received2016-02-16
Model Number10.0.2
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMERGE HEALTHCARE
Manufacturer Address900 WALNUT RIDGE DRIVE HARTLAND WI 53029 US 53029


Patients

Patient NumberTreatmentOutcomeDate
10 2016-02-16