MERGE EYE STATION

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2016-02-16 for MERGE EYE STATION manufactured by Merge Healthcare.

Event Text Entries

[38227887] Upon restrospective review, this issue was determined to be reportable as an mdr.
Patient Sequence No: 1, Text Type: N, H10


[38227888] Merge eye station is an image capture software and digital interface for use in conjunction with existing opthalic fundus cameras to take images of the eye. It performs fluorescein angiography, ref free and color, icg still-image photography and video imaging. On 7/20/15, merge was contacted to report that images were imported under "unknown". Merge took out l for comparepatient_db in esiu on all import stations. The customer was shown how to complete this and the system was restarted. The systems are design to allow sites to validate patient information at the time of import on one patient's demographics. Additionally, certain releases will overwrite the patient name if the medical record numbers (mrns) match but the names do not or will import the patient data as "unknown". Validation on one patient's demographics could potentially introduce patient data mixup or delay in patient care. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183926-2015-00319
MDR Report Key5437932
Report SourceHEALTH PROFESSIONAL
Date Received2016-02-16
Date of Report2016-02-15
Date of Event2015-07-20
Date Mfgr Received2015-07-20
Date Added to Maude2016-02-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHAEL DIEDRICK
Manufacturer Street900 WALNUT RIDGE DR.
Manufacturer CityHARTLAND WI 53029
Manufacturer CountryUS
Manufacturer Postal53029
Manufacturer Phone2629123570
Manufacturer G1MERGE HEALTHCARE
Manufacturer Street900 WALNUT RIDGE DRIVE
Manufacturer CityHARTLAND WI 53029
Manufacturer CountryUS
Manufacturer Postal Code53029
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number2183926-02/15/2016-039-C
Event Type3
Type of Report3

Device Details

Brand NameMERGE EYE STATION
Generic NameSYSTEM, IMAGE MANAGEMENT, OPHTHALMIC
Product CodeNFJ
Date Received2016-02-16
Model NumberMERGE EYE STATION
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMERGE HEALTHCARE
Manufacturer Address900 WALNUT RIDGE DRIVE HARTLAND WI 53029 US 53029


Patients

Patient NumberTreatmentOutcomeDate
10 2016-02-16

© 2026 FDA.report