FREE PSA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-02-17 for FREE PSA manufactured by Roche Diagnostics.

Event Text Entries

[38345931] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[38345932] The customer reported that they received questionable results for a total of 50 patient samples tested for total prostate-specific antigen (tpsa) and free prostate-specific antigen (fpsa) tested on an e601 analyzer. Data was provided for a total of three patient samples and of these three samples, two had erroneous results for tpsa and fpsa. The fpsa results were higher than the tpsa results. It was asked, but it is not known if any erroneous results were reported outside of the laboratory. The units of measure used for the tpsa and fpsa assays were asked for, but not provided. This medwatch will cover fpsa. Please refer to the medwatch with patient identifier (b)(6) for information related to tpsa. The first sample resulted as 0. 118 for fpsa and 0. 005 for tpsa on (b)(6) 2015. The second sample resulted as 0. 118 for fpsa and 0. 010 for tpsa on (b)(6) 2015. The patients were not adversely affected. The e601 analyzer serial number was (b)(4). A specific root cause could not be determined based on the provided information. The patient samples were requested for investigation, but could not be provided.
Patient Sequence No: 1, Text Type: D, B5


[39108862] The units of measure used for the tpsa and fpsa assays are ng/ml.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2016-00172
MDR Report Key5441201
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-02-17
Date of Report2016-02-26
Date of Event2015-12-15
Date Mfgr Received2016-01-26
Date Added to Maude2016-02-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2016-02-17
Model NumberNA
Catalog NumberASKU
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2016-02-17