HARMONY IQ 2800 INTEGRATION SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-02-19 for HARMONY IQ 2800 INTEGRATION SYSTEM manufactured by Vts Medical Systems.

Event Text Entries

[39133929] A steris service technician inspected the integration system and found that the touch panel was not responding. The touch panel present during the time of the reported event was a third party touch panel. The harmony iq 2800 integration system has usb ports which allow for the user to enable use of the integration system should the touch panel not operate properly. Although the user facility reported an isssue with their touch panel, the harmony iq 2800 integration was still active and able to be utilized via other manual override methods. The technician replaced the touch panel subject of the reported event with a steris touch panel, tested the system and confirmed it to be operational. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[39133930] The user facility reported that their touch panel for their integration system was not operating properly. No report of injury however, procedural delays were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1000404456-2016-00001
MDR Report Key5447157
Date Received2016-02-19
Date of Report2016-02-19
Date of Event2016-01-22
Date Mfgr Received2016-01-22
Date Added to Maude2016-02-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1VTS MEDICAL SYSTEMS
Manufacturer Street40 MELVILLE PARK RD
Manufacturer CityMELVILLE NY 11747
Manufacturer CountryUS
Manufacturer Postal Code11747
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHARMONY IQ 2800 INTEGRATION SYSTEM
Generic NameINTEGRATION SYSTEM
Product CodeKQM
Date Received2016-02-19
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVTS MEDICAL SYSTEMS
Manufacturer Address40 MELVILLE PARK RD MELVILLE NY 11747 US 11747


Patients

Patient NumberTreatmentOutcomeDate
10 2016-02-19