TEC 6 PLUS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-02-23 for TEC 6 PLUS manufactured by Datex-ohmeda, Inc..

Event Text Entries

[39303897] Ge is awaiting the return of the malfunctioning vaporizer before beginning an evaluation. A follow-up report will be submitted upon completion of the evaluation. No report of patient involvement. Device evaluation anticipated, but not yet begun.
Patient Sequence No: 1, Text Type: N, H10


[39303898] Third party service organization reported that, during routine maintenance, it was noted that the unit's interlock system was not functioning. There was no report of patient involvement.
Patient Sequence No: 1, Text Type: D, B5


[43555009] The unit was returned to the manufacturing site for investigation. The unit was tested and the reported complaint was confirmed. During inspection of the unit, it was noted that the interlock plunger pin circlip was missing. The circlip was replaced, resolving the reported complaint.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2112667-2016-00336
MDR Report Key5454022
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-02-23
Date of Report2016-04-25
Date of Event2016-01-25
Date Mfgr Received2016-04-14
Device Manufacturer Date1970-01-01
Date Added to Maude2016-02-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBRAD WHEELER
Manufacturer Street3000 N. GRANDVIEW BLVD., W450
Manufacturer CityWAUKESHA WI 53188
Manufacturer CountryUS
Manufacturer Postal53188
Manufacturer G1DATEX-OHMEDA, INC.
Manufacturer Street3030 OHMEDA DRIVE
Manufacturer CityMADISON WI 53718
Manufacturer CountryUS
Manufacturer Postal Code53718
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTEC 6 PLUS
Generic NameVAPORIZER
Product CodeCAD
Date Received2016-02-23
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDATEX-OHMEDA, INC.
Manufacturer Address3030 OHMEDA DRIVE MADISON WI 53718 US 53718


Patients

Patient NumberTreatmentOutcomeDate
10 2016-02-23

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.