THINPREP IMAGING SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-02-26 for THINPREP IMAGING SYSTEM manufactured by Hologic, Inc..

Event Text Entries

[39231777] Possible false negative. Per hologic's cytology applications specialist (cas) a customer's lab personnel reported a case with observed abnormal cells that were not located in the 22 fovs selected by the imager. Cas reviewed slide on the review scope as an unknown mixed with other slides. Case presented had no triggers in the 22 fovs to prompt an autoscan. Full review of slide during qc showed abnormal cells outside. Case was picked up on qc. There were two groups of hsil cells on the slide, with no abnormal cells in the 22 fov. The customer would like to submit the slide for internal review though the cas explained they would not receive a report. This is a reportable event in the us as this is a potential delay in diagnosis.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1222780-2016-00043
MDR Report Key5463602
Date Received2016-02-26
Date of Report2016-02-25
Date Mfgr Received2016-02-03
Date Added to Maude2016-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS EVA MAXWELL
Manufacturer Street250 CAMPUS DRIVE
Manufacturer CityMARLBOROUGH, MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTHINPREP IMAGING SYSTEM
Generic NameAUTOMATED MICROSCOPE FOR CYTOLOGY
Product CodeMNM
Date Received2016-02-26
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHOLOGIC, INC.
Manufacturer Address250 CAMPUS DRIVE MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-02-26