MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign,health pr report with the FDA on 2016-02-26 for (COC/AMP/EDDP100/MOP/BZO)MULTI-DRUG ONE STEP 5 DRUG SCREEN TEST PANEL (URINE) DOA-154-1661-XA-ONC536B manufactured by Alere San Diego, Inc..

Event Text Entries

[39209899] Investigation conclusion: customer's observation was not replicated in-house with retention devices. Retention devices were tested with in-house drug free donor urine; all coc results were negative at 5 min read time. No false positive results were obtained. Manufacturing batch record review did not uncover any abnormalities. Root cause could not be determined from the information provided. Based on the information available, there is no indication of a product deficiency. No corrective action is required at this time.
Patient Sequence No: 1, Text Type: N, H10


[39209900] Customer alleging receiving false positive coc, alternative method negative results. Event occurred in (b)(6). Alternate testing method not provided. Coc was compared against another rapid test from another company, which showed negative results. No reported adverse patient sequela.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2016-00120
MDR Report Key5464681
Report SourceDISTRIBUTOR,FOREIGN,HEALTH PR
Date Received2016-02-26
Date of Report2016-02-02
Date of Event2016-01-01
Date Mfgr Received2016-02-02
Date Added to Maude2016-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name(COC/AMP/EDDP100/MOP/BZO)MULTI-DRUG ONE STEP 5 DRUG SCREEN TEST PANEL (URINE)
Generic NameDRUG OF ABUSE TEST
Product CodeDKZ
Date Received2016-02-26
Model NumberDOA-154-1661-XA-ONC536B
Lot NumberDN-DOA5070775
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2016-02-26