TRIGLYCERIDES_2 REAGENT, CONCENTRATED 10697575

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-03-10 for TRIGLYCERIDES_2 REAGENT, CONCENTRATED 10697575 manufactured by Siemens Healthcare Diagnostics, Inc..

Event Text Entries

[40136491] Siemens healthcare diagnostics has confirmed the trig_c reagent kit lots 348297 and 359932( smn (b)(4)) used on the advia? 1200, 1650, 1800, 2400, and xpt chemistry systems does not meet instructions for use (ifu) linearity claim at the upper limit of the assay range as it approaches end of shelf life. Internal testing demonstrated that a patient bias (-12% to -15%) may be seen at concentration levels between 450 to 550mg/dl. Linearity is reduced by approximately 31% at higher triglyceride concentrations up to 1200mg/dl (13. 56mmol/l). An urgent field safety notice (ufsn) chc16-03. A. Ous was sent to ous customers and a urgent medical device recall (umdr) chc16-03. A. Us was sent to us customers in march of 2016. The ufsn and umdr state that customers are to discontinue use and discard reagent kit lots 348297 and 359932.
Patient Sequence No: 1, Text Type: N, H10


[40136492] The customer has noticed a positive bias in results on patient samples for the triglycerides_2 concentrated assay when using reagent lot 348297 on the advia chemistry 1800 instrument. The initial result obtained on a patient sample was low. The low result was reported and questioned by the physician(s). The patient sample was then diluted and the results was higher. The dilution was repeated and the result was comparable to the first result on the same dilution. The second result on the dilution was reported to the physician(s). Another patient sample with a high trig_c value was also diluted and the result was higher. Unknown if either the initial or repeat result was reported to the physician(s). The customer then tested more patient samples at two different dilutions and observed a positive bias on the all the results. There were no reports of patient intervention or adverse health consequences due to the trig_c results obtained on the patient samples.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2432235-2016-00112
MDR Report Key5492108
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-03-10
Date of Report2016-02-29
Date of Event2016-02-14
Date Mfgr Received2016-02-29
Date Added to Maude2016-03-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactAARTI AZIZ
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242683
Manufacturer G1RANDOX LABORATORIES LTD (REG # 8020890)
Manufacturer Street55 DIAMON ROAD
Manufacturer CityCRUMLIN, CO. ANTRIM BT294QY
Manufacturer CountryUK
Manufacturer Postal CodeBT29 4QY
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number2432235-03/09/2016-001-R
Event Type3
Type of Report3

Device Details

Brand NameTRIGLYCERIDES_2 REAGENT, CONCENTRATED
Generic NameTRIGLYCERIDES_2 REAGENT, CONCENTRATED
Product CodeCDT
Date Received2016-03-10
Model NumberTRIGLYCERIDES_2 REAGENT, CONCENTRATED
Catalog Number10697575
Lot Number348297
Device Availability*
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS, INC.
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.