TRIGLYCERIDES_2 REAGENT, CONCENTRATED 10697575

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-03-10 for TRIGLYCERIDES_2 REAGENT, CONCENTRATED 10697575 manufactured by Siemens Healthcare Diagnostics, Inc..

Event Text Entries

[40134037] Siemens healthcare diagnostics has confirmed the trig_c reagent kit lots 348297 and 359932( smn (b)(4)) used on the advia? 1200, 1650, 1800, 2400, and xpt chemistry systems does not meet instructions for use (ifu) linearity claim at the upper limit of the assay range as it approaches end of shelf life. Internal testing demonstrated that a patient bias (-12% to -15%) may be seen at concentration levels between 450 to 550mg/dl. Linearity is reduced by approximately 31% at higher triglyceride concentrations up to 1200mg/dl (13. 56mmol/l). An urgent field safety notice (ufsn) (b)(4)ous was sent to ous customers and a urgent medical device recall (umdr) chc16-03. A. Us was sent to us customers in (b)(6) 2016. The ufsn and umdr state that customers are to discontinue use and discard reagent kit lots 348297 and 359932.
Patient Sequence No: 1, Text Type: N, H10


[40134038] The customer has observed that the linearity claims for the triglycerides_2 concentrated assay are not being met when using reagent lot 348297 on the advia chemistry 2400 instrument. High patient samples when diluted report results higher than the result obtained on a sample when run neat. The high result obtained on the diluted patient sample matches with the expected result. It is unknown if the results were reported to the physician(s). There were no reports of patient intervention or adverse health consequences due to the trig_c assay not meeting linearity claims.
Patient Sequence No: 1, Text Type: D, B5


[40549214] The initial mdr 2432235-2016-00124 was filed on 3/10/2016. Corrected information: in the initial mdr the date of event was indicated as (b)(6) 2016. The customer has informed siemens diagnostics on (b)(6) 2016, that the correct date of event is (b)(6) 2016. Additional information (3/16/2016): in the initial mdr it was unknown if the patient results were reported to the physician(s). The customer has informed siemens healthcare that results above 400 ng/dl were not reported to physicians(s). The customer confirms all samples >400 ng/d with dilutions.
Patient Sequence No: 1, Text Type: N, H10


[40549215] Results above 400 ng/dl were not reported to physicians(s). The customer confirms all samples >400 ng/d with dilutions.
Patient Sequence No: 1, Text Type: D, B5


[41977882] The initial mdr 2432235-2016-00124 was filed on 3/10/2016. The dates were incorrectly listed as (b)(6) 2016 and 03/02/2016 in the event date and report date fields respectively. The correct dates are 03/02/2016 for report date and 02/29/2016 for date received by mfr.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2016-00124
MDR Report Key5492114
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-03-10
Date of Report2016-03-02
Date of Event2016-03-01
Date Mfgr Received2016-03-16
Date Added to Maude2016-03-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactAARTI AZIZ
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242683
Manufacturer G1RANDOX LABORATORIES LTD. (REG# 8020890)
Manufacturer Street55 DIAMON ROAD CRUMLIN
Manufacturer CityCO. ANTRIM, BT294QY
Manufacturer CountryUK
Manufacturer Postal CodeBT29 4QY
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number2432235-03/09/2016-001-R
Event Type3
Type of Report3

Device Details

Brand NameTRIGLYCERIDES_2 REAGENT, CONCENTRATED
Generic NameTRIGLYCERIDES_2 REAGENT, CONCENTRATED
Product CodeCDT
Date Received2016-03-10
Model NumberTRIGLYCERIDES_2 REAGENT, CONCENTRATED
Catalog Number10697575
Lot Number348297
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS, INC.