RELIANCE ENDOSCOPE PROCESSING SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-03-10 for RELIANCE ENDOSCOPE PROCESSING SYSTEM manufactured by Steris Canada Corporation.

Event Text Entries

[40325819] A steris field service technician arrived onsite, inspected the unit, and identified a cut on the unit's door gasket. Water was able to pass through the gasket. It is unknown how the unit sustained the reported damage. The technician replaced the door gasket, tested the unit, and confirmed it to be operating according to specification. The unit was returned to service. Section 1 of the operator manual for the reliance eps states, "to avoid slippery floor conditions, immediately wipe up any spilled liquids or drippage. " no additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[40325820] The user facility reported their reliance endoscope processing system was leaking water. No report of injury or procedural delay or cancellation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9680353-2016-00017
MDR Report Key5493381
Date Received2016-03-10
Date of Report2016-03-10
Date of Event2016-02-12
Date Mfgr Received2016-02-12
Date Added to Maude2016-03-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC 8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameRELIANCE ENDOSCOPE PROCESSING SYSTEM
Generic NameENDOSCOPE PROCESSOR
Product CodeNZA
Date Received2016-03-10
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC 8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.