MICROTARGETING ELECTRODE FC2002

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2016-03-11 for MICROTARGETING ELECTRODE FC2002 manufactured by Fhc, Inc,.

Event Text Entries

[40137878] Product catalog number fc2002, microtargeting electrode, lot # 212327. Event date was (b)(6) 2016 at (b)(6). Event was described as follows: "during the first (left) side of the stage 1 there was satisfactory mer but then some unusual side effects (details unknown) experienced during the test stimulation. The neurologist and neurosurgeon decided to take a different trajectory. The second pass was 2mm anterolateral using the standard star drive array spacing. Mer was seemingly quiet down most of the tract. The neurologist decided to test stimulate at target but before any stimulation was delivered the neurologist decided to abandon the case due to the patient's condition and inability to speak. The fc2002 was removed and the neurosurgeon closed the burr hole and skin incision. The frame was removed and the patient had a ct scan, which showed a bleed in the basal ganglia region of the brain. The case was abandoned. 1-2 hours after the ct scan was performed it was reported that the patient was improving and was medically stable. " the electrode was discarded and thus will not be returned for analysis. Patient was improving and determined medically stable 1-2 hours after the ct scan was performed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002250546-2016-00001
MDR Report Key5494583
Report SourceDISTRIBUTOR,FOREIGN
Date Received2016-03-11
Date of Report2016-03-02
Date of Event2016-02-12
Date Mfgr Received2016-02-15
Date Added to Maude2016-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CRAIG PAGAN
Manufacturer Street1201 MAIN STREET
Manufacturer CityBOWDOIN ME 04287
Manufacturer CountryUS
Manufacturer Postal04287
Manufacturer Phone2076665651
Manufacturer G1FHC, INC,
Manufacturer Street1201 MAIN STREET
Manufacturer CityBOWDOIN ME 04287
Manufacturer CountryUS
Manufacturer Postal Code04287
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMICROTARGETING ELECTRODE
Generic NameELECTRODE, DEPTH
Product CodeGZL
Date Received2016-03-11
Model NumberFC2002
Lot Number212327
Device Expiration Date2018-03-15
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFHC, INC,
Manufacturer Address1201 MAIN STREET BOWDOIN ME 04287 US 04287


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2016-03-11