TPS MICRO DRIVER 5100088000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-03-11 for TPS MICRO DRIVER 5100088000 manufactured by Stryker Instruments-kalamazoo.

Event Text Entries

[40183340] The failure for oscillating in forward and reverse was duplicated through functional testing. Disassembly and visual inspection by the technician noted the power control board (pcb)/socket was corroded. Corrosion may limit the design properties of the device causing the event. The pcb assembly was replaced.
Patient Sequence No: 1, Text Type: N, H10


[40183341] It was reported by the user facility that during a procedure the device oscillated in forward and reverse continually. The procedure was completed successfully with no delay, no medical intervention, and no adverse consequences reported.
Patient Sequence No: 1, Text Type: D, B5


[42276810]
Patient Sequence No: 1, Text Type: N, H10


[42276811] It was reported by the user facility that during a procedure the device oscillated in forward and reverse continually. The procedure was completed successfully with no delay, no medical intervention, and no adverse consequences reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001811755-2016-00288
MDR Report Key5495917
Report SourceUSER FACILITY
Date Received2016-03-11
Date of Report2016-02-15
Date of Event2016-02-15
Date Mfgr Received2016-02-15
Device Manufacturer Date1999-04-20
Date Added to Maude2016-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CASEY METZGER
Manufacturer Street4100 EAST MILHAM AVENUE
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal49001
Manufacturer Phone2693237700
Manufacturer G1STRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Street4100 EAST MILHAM AVENUE
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal Code49001
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTPS MICRO DRIVER
Generic NameDRIVER, WIRE, AND BONE DRILL, MANUAL
Product CodeDZJ
Date Received2016-03-11
Returned To Mfg2016-02-19
Catalog Number5100088000
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Address4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.