ETEST? ITRACONAZOLE 525818

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-03-14 for ETEST? ITRACONAZOLE 525818 manufactured by Biomerieux, Sa.

Event Text Entries

[40275478] A customer in (b)(6) reported having issues with the mic on qc while using; etest itraconazole it 32 ww f100. The results are as follows; both c. Albicans (atcc 90028) and c. Parapsilosis get an mic of 0. 032. There is no evidence/indication of the following: injury or death to a patient, evidence that a patient underwent any unnecessary medical procedure due to the reported event, physician alleging the reported event caused him/her to provide incorrect treatment.
Patient Sequence No: 1, Text Type: D, B5


[47816746] An internal investigation was initiated due to the report of low mic results with etest itraconazole and two (2) atcc strains: candida parapsilosis atcc 22019 and candida albicans atcc 90028. Results of the internal investigation are as follows: since 2010, there has been only one (1) complaint about this molecule. Inoculation by floating is not within the recommendations of intended use and it is also necessary for antifungals to swabbed twice. Recommendation of package insert: soak a sterile, non-toxic and not too tightly spun swab in the inoculum suspension and remove excess fluid by pressing it against the inside wall of the test tube. Carefully streak the entire agar surface three times, rotating the plate 60 degrees each time to evenly distribute the inoculum. Soak the swab again and repeat the streaking procedure a second time. Alternatively, use retro c80 (rota-plater, biom? Rieux) to efficiently streak the inoculum over the agar surface and then soak the swab again before streaking a second time. Allow excess moisture to be absorbed for approximately 15 to 20 minutes so that the surface is completely dry before applying the etest gradient strips. The conformity of this batch was confirmed with three (3) atcc stains (atcc 90028 - atcc 6258 - atcc 22019). Testing was performed on the customer's strips and the retained samples corresponding to the batch 1003467650 in parallel with another batch as reference. Result- unable to reproduce the customer's observation with the quality control strains.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2016-00053
MDR Report Key5498174
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-03-14
Date of Report2016-03-04
Date Mfgr Received2016-03-04
Device Manufacturer Date2014-09-03
Date Added to Maude2016-03-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME LES GROTTES ISERE, 38390
Manufacturer CountryFR
Manufacturer Postal Code38390
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? ITRACONAZOLE
Generic NameETEST? ITRACONAZOLE
Product CodeNGZ
Date Received2016-03-14
Catalog Number525818
Lot Number1003467650
Device Expiration Date2017-09-29
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE, 38390 FR 38390


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.